Series II Standard Mass Flow Controller | Product Series

필터 기준:

    선택한 필터 없음

지금 Parker 계정을 만들어 보십시오!

계정을 만들면 구매 선호사항에서 애플리케이션 액세스에 이르는 모든 Parker 관련 정보를 관리할 수 있습니다.


LOADING IMAGES
Series II Standard Mass Flow Controller

Series II Standard Mass Flow Controller

Our Series II mass flow controllers & meters offer outstanding value in a wide range of applications. Exceptional accuracy for flow ranges from 0.014 sccm to 1,670 slpm, all with optional multi-gas and multi-range capability for maximum flexibility.

기술 사양

  • 유량: Up to 1670 L/min
  • 압력 제어 범위: 928 PSIA - 1450 PSIA
  • 연결 유형: Compression, Face Seal Male, None, Other
  • 연결 크기: 1/8"-1/2" | 6 mm - 20 mm
  • 통신 인터페이스: Analog or digital: RS-232, DeviceNet, Profibus-DP, Modbus-RTU
  • 밸브 유형: Normally Open, Normally Closed
  • 밀폐재 재료: EPDM, Kalrez, Viton (Standard)
  • 바디 재료: 316L Stainless Steel
  • 유량 정확성: Standard: ± 0.5% Reading plus ± 0.1% Full Scale (± 1% Full Scale for ranges 3 - 5 ml/min; ± 2% Full Scale for ranges < 3 ml/min)
  • 턴다운비: 1 : 50 (in digital mode up to 1 : 187.5)
  • 반복성: < 0.2% reading
  • 정착 시간: Standard: 2 - 4 seconds
  • 제어 안정성: < ±0.1% Full Scale (typical for 1 l/min N2)
  • 최고 작동 온도: 70 °C
  • 최저 작동 온도: -10 °C
  • 온도 계수: Zero: < 0.05% Full Scale / °C; span: < 0.05% Reading / °C °C
  • 압력 계수: 0.1 ATM typical N2; 0.01% / ATM typical H2 ATM
  • 누설률: Tested < 2 x 10-9 mbar l/s He L/sec
  • 자세 민감도: Max. error at 90° off horizontal 0.2% at 1 ATM, typical N2
  • 예열 시간: 30 min. for optimum accuracy 2 min. for accuracy ±2% Full Scale
  • 습식 재료: 316L Stainless Steel
  • 표면 마감: 0.8 µm typical
  • 연결 재료: 316L Stainless Steel
  • 충족된 사양: IP40
  • 전원공급장치 전압: +15 - 24 Vdc
  • 소비 전력: Max. 320 mA; Add 50 mA Profibus, if applicable
  • 전기 연결: 9-PIN D-Connector

전체 제품 설명

Parker's highly precise mass flow controllers provide you the diagnostic stability and repeatability you need for low flow gas applications. Controlling thermal mass flow rates from 0.6 l/m to 1,670 l/m with high precision and <0.2% repeatability.  Rated for 8 separate gas types which are customer field adjustable. Our thermal mass flow controller is available in both digital and analog styles, fully compatible with RS232 and digital protocols such as Profibus.  If you're not sure what gas you're going to use or your process changes, our MFC allows you to migrate to a new gas within your flow range without purchasing a new mass flow controller.

Markets:
• Laboratory and Process Instrumentation
• Pharma/Bio-Pharma Equipment
• Air Quality Monitoring System
• Furnace and Coatings

Features/Benefits:
• Fast, Repeatable and Reliable Performance
• Premium Accuracy with Proven Thermal Mass Flow Sensor and Laminar Flow Element
• Easy to integrate into your new or existing systems
• Asset calibration management software included
• Cleaned for analytical use
• Fails shut with Normally Closed Valve for safe operation
• CE, REACH and RoHS II

Applications:
• Gas Control for Laboratory and Process Equipment
• Burner Ratio Control for Ceramics and Metals
• Process and Environmental Analyzers
• Emissions Monitors and Calibrators

관련 문서

×

687611